CVE-2017-9657 (GCVE-0-2017-9657)
Vulnerability from cvelistv5 – Published: 2018-04-30 15:00 – Updated: 2024-09-16 16:13
VLAI?
Summary
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Severity ?
No CVSS data available.
CWE
- CWE-460 - Improper cleanup on thrown exception CWE-460
Assigner
References
| URL | Tags | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
|
|||||||||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Philips | IntelliVue MX40 Patient Worn Monitor |
Affected:
IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18
|
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-05T17:18:01.347Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM",
"x_transferred"
],
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"name": "100813",
"tags": [
"vdb-entry",
"x_refsource_BID",
"x_transferred"
],
"url": "http://www.securityfocus.com/bid/100813"
},
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "IntelliVue MX40 Patient Worn Monitor",
"vendor": "Philips",
"versions": [
{
"status": "affected",
"version": "IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18"
}
]
}
],
"datePublic": "2017-09-12T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-460",
"description": "Improper cleanup on thrown exception CWE-460",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2018-05-01T09:57:01.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM"
],
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"name": "100813",
"tags": [
"vdb-entry",
"x_refsource_BID"
],
"url": "http://www.securityfocus.com/bid/100813"
},
{
"tags": [
"x_refsource_MISC"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2017-09-12T00:00:00",
"ID": "CVE-2017-9657",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "IntelliVue MX40 Patient Worn Monitor",
"version": {
"version_data": [
{
"version_value": "IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18"
}
]
}
}
]
},
"vendor_name": "Philips"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Improper cleanup on thrown exception CWE-460"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.usa.philips.com/healthcare/about/customer-support/product-security",
"refsource": "CONFIRM",
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"name": "100813",
"refsource": "BID",
"url": "http://www.securityfocus.com/bid/100813"
},
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2017-9657",
"datePublished": "2018-04-30T15:00:00.000Z",
"dateReserved": "2017-06-14T00:00:00.000Z",
"dateUpdated": "2024-09-16T16:13:21.087Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…