CVE-2024-9832 (GCVE-0-2024-9832)
Vulnerability from cvelistv5 – Published: 2024-11-14 21:03 – Updated: 2024-11-18 15:37
VLAI?
Title
No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
Summary
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Severity ?
9.3 (Critical)
CWE
- CWE-307 - Improper Restriction of Excessive Authentication Attempts
Assigner
References
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Baxter | Life2000 Ventilation System |
Affected:
06.08.00.00 and prior
|
{
"containers": {
"adp": [
{
"affected": [
{
"cpes": [
"cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"
],
"defaultStatus": "unaffected",
"product": "life2000_ventilator_firmware",
"vendor": "baxter",
"versions": [
{
"lessThanOrEqual": "06.08.00.00",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
],
"metrics": [
{
"other": {
"content": {
"id": "CVE-2024-9832",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2024-11-18T15:36:22.221928Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2024-11-18T15:37:00.311Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "Life2000 Ventilation System",
"vendor": "Baxter",
"versions": [
{
"status": "affected",
"version": "06.08.00.00 and prior"
}
]
}
],
"datePublic": "2024-11-14T21:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.\u003cbr\u003e"
}
],
"value": "There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure."
}
],
"impacts": [
{
"capecId": "CAPEC-49",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-49 Password Brute Forcing"
}
]
},
{
"capecId": "CAPEC-441",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-441 Malicious Logic Insertion"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 9.3,
"baseSeverity": "CRITICAL",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "CHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-307",
"description": "CWE-307 Improper Restriction of Excessive Authentication Attempts",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2024-11-14T21:47:11.069Z",
"orgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
"shortName": "Baxter"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"
}
],
"source": {
"discovery": "UNKNOWN"
},
"title": "No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
}
},
"cveMetadata": {
"assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
"assignerShortName": "Baxter",
"cveId": "CVE-2024-9832",
"datePublished": "2024-11-14T21:03:16.721Z",
"dateReserved": "2024-10-10T19:24:46.919Z",
"dateUpdated": "2024-11-18T15:37:00.311Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1",
"vulnerability-lookup:meta": {
"vulnrichment": {
"containers": "{\"adp\": [{\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2024-9832\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"none\"}, {\"Automatable\": \"no\"}, {\"Technical Impact\": \"total\"}], \"version\": \"2.0.3\", \"timestamp\": \"2024-11-18T15:36:22.221928Z\"}}}], \"affected\": [{\"cpes\": [\"cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*\"], \"vendor\": \"baxter\", \"product\": \"life2000_ventilator_firmware\", \"versions\": [{\"status\": \"affected\", \"version\": \"0\", \"versionType\": \"custom\", \"lessThanOrEqual\": \"06.08.00.00\"}], \"defaultStatus\": \"unaffected\"}], \"providerMetadata\": {\"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"shortName\": \"CISA-ADP\", \"dateUpdated\": \"2024-11-18T15:36:55.272Z\"}}], \"cna\": {\"title\": \"No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator\", \"source\": {\"discovery\": \"UNKNOWN\"}, \"impacts\": [{\"capecId\": \"CAPEC-49\", \"descriptions\": [{\"lang\": \"en\", \"value\": \"CAPEC-49 Password Brute Forcing\"}]}, {\"capecId\": \"CAPEC-441\", \"descriptions\": [{\"lang\": \"en\", \"value\": \"CAPEC-441 Malicious Logic Insertion\"}]}], \"metrics\": [{\"format\": \"CVSS\", \"cvssV3_1\": {\"scope\": \"CHANGED\", \"version\": \"3.1\", \"baseScore\": 9.3, \"attackVector\": \"LOCAL\", \"baseSeverity\": \"CRITICAL\", \"vectorString\": \"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\", \"integrityImpact\": \"HIGH\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"HIGH\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Baxter\", \"product\": \"Life2000 Ventilation System\", \"versions\": [{\"status\": \"affected\", \"version\": \"06.08.00.00 and prior\"}], \"defaultStatus\": \"unaffected\"}], \"datePublic\": \"2024-11-14T21:00:00.000Z\", \"references\": [{\"url\": \"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\"}], \"x_generator\": {\"engine\": \"Vulnogram 0.2.0\"}, \"descriptions\": [{\"lang\": \"en\", \"value\": \"There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.\u003cbr\u003e\", \"base64\": false}]}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-307\", \"description\": \"CWE-307 Improper Restriction of Excessive Authentication Attempts\"}]}], \"providerMetadata\": {\"orgId\": \"dba971b9-eb30-4121-91e1-3b45611354aa\", \"shortName\": \"Baxter\", \"dateUpdated\": \"2024-11-14T21:47:11.069Z\"}}}",
"cveMetadata": "{\"cveId\": \"CVE-2024-9832\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2024-11-18T15:37:00.311Z\", \"dateReserved\": \"2024-10-10T19:24:46.919Z\", \"assignerOrgId\": \"dba971b9-eb30-4121-91e1-3b45611354aa\", \"datePublished\": \"2024-11-14T21:03:16.721Z\", \"assignerShortName\": \"Baxter\"}",
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
}
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…